Performance Specification And Tolerances
Ventilator
Parameters Ranges Tolerances Controls Power ON/OFF Modes
C
A/C
SIMV Tidal Volume 200~1200ml ¡À20% Breath Rate 5~60 bpm +/-20% Sigh 1~10 per 100 breath Trigger Level -10~0hpa +/-1hpa Over Pressure Relief ¡Ü60hpa
Performance Specification And Tolerances
Ventilator
Parameters Ranges Tolerances Alarms and Indicators High Pressure Alarm Limit 2~6kpa
flash, red indicator +/-0.1kpa Low Pressure Alarm Limit 0. 2~2kpa
Red indicator +/-0.1kpa Alarm Silence 60 sec. max
Performance Specification And Tolerances Ventilator
Parameters Ranges Tolerances Monitors Tidal Volume 0~1500ml +/-20% Total Breath Rate 5~60 bpm +/-20% Main Circuit Pressure -1~6KPa +/-20% Wave Display Main Circuit Pressure
What Is Provided Parts Number Specification Ventilator 1 Oxygen bottle 1 3L .can be used for 1.5 hour Take Box 1 all parts in it Breath Circuit 1 1.5m Breath value 1 Mast 1 Charger 1 Battery 1 Can be used for 8 hours Wrench 1
|